GPS

DGC-T TABLETOP PULSE OXIMETER

★. Three Parameters: SPO2, Pulse Rate, TEMP (Optional)

★. Waveform: SPO2 waveform (PLETH)

Features:

★. Compact, small, light, easy for carrying and handling;

★. LED display of measurement and LCD display of PLETH;

★. Audio and visual alarm;

★. Storage and review of patient’s trend data up to 36 hours which is never lost even when power’s off;

★. Three kinds of power supply, AC Power, Build-in rechargeable Ni/MH battery and Common AA alkaline or rechargeable batteries;

★. Battery capacity indicator;

★. Automatic power-off;

★. Patient’s trend can be transmitted to a personal computer for display,saving and printing;

★. Suitable for adult, pediatric, and neonate patients; 

Applications:

Table Pulse Oximeter can be applied in the hospital’s operation room, ICU, clinic section office, Out-patient department, sickroom and emergency treatment.

Standard Packing List:

Tabletop Pulse Oximeter main unit 1pc

Adult finger SpO2 sensor 1pc

AC power adaptor 1pc

Communication cable 1pc

The operator’s manual 1pc 

Selectable Accessory:

Surface TEMP probe

Child finger/toe SpO2 sensor  

H.S.Code: 90189090  

Dimension: 160x148x85mm  

Net Weight: 0.5kg   

Gross Weight: 1kg  

Package Size: 245x185x100mm

Color: Light blue

Technical Specification:

(1) General

Monitoring Parameters: SPO2、PR、PLETH、Pulse strength、TEMP(Optional)

 Signal sockets: SPO2 socket、PR socket、TEMP socket(Optional)、adaptor socket、PC communication socket

Display Screen        

Type: 128×64 dot matrix homochromy LCD and Dual-color LED (red/green)

LCD Display area: 46mm×32mm

LED Display area: 35mm×58mm

Backlight color: White

LED: Dual-color LED (red/green) and homochromy LED

(2) Features

PLETH: The monitor can display PLETH waveform

 Audio and visual alarm: The speaker sounds the alarm and the parameter’s measurement flashes on the screen.

 Prompting: Icons of battery capacity, pulse sound switch and alarm sound switch.

 Power saving: The monitor will automatic shut down when no monitoring for 3 minutes.

Trend: The monitor saves parameter measurements every 2 seconds and can store trend data up to 36 hours.

Data storage: The saved data will never be lost even if the monitor is powered off.

Communication: Transmitting the patient’s trend to a PC for display, observation, saving and printing. 

(3) Electrical Specifications

Working Voltage: DC 6.0V~9.0V

AC power adaptor      

Input: AC 100~240V, 50/60Hz, 0.7A

 Output: DC 5V,3A

Internal battery   

Type: 1.5Ahr/8.4V Li rechargeable battery

Run Time: 15-hour continuous operation with a new, fully charged battery (environment temperature is 25℃)

 Recharge time: 5 hours

Common 1.5V Battery

 Type: Five pieces of common 1.5V AA battery or rechargeable battery

 Run Time: More than 5-hour continuous operation with a new battery (environment temperature is 25℃) 

(4) Environment

Temperature

Operation: 0°C ~ 45 °C 

Transportation and storage: -20 °C ~ 60 °C

Humidity

Operation: 15% ~95 %(non-condensing)  

Transportation and storage: 10% ~ 95 % (non-condensing)  

Altitude

Operation: 86KPa~106KPa

Transportation and storage: 50KPa~106KPa 

(5) Parameter Specifications

SPO2

Patient: Adult, child and neonate

Range: 35%~100%

Resolution: 1 %

Accuracy: ±2 %(70% ~100%), Unspecified (0%~ 69%)

PR 

Range: 30 bpm~250bpm  

Resolution: 1bpm

Accuracy: ±2bpm

TEMP(Optional)

Channel: 1  

Input: Body surface thermal-sensitive resistor temperature sensor  

Range: 0℃~50℃  

Resolution: 0.1℃

Accuracy: ±0.2℃   