GPS

DGC-II(AA) HANDHELD PULSE OXIMETER

★. Two Parameter: SPO2,Pulse Rate

Function:

★. Measuring: Display of SPO2, PR and pulse strength.

★. Prompting: Battery capacity.

★. Alarm: Judgment of exceeding the limits, Audio and visual alarm.

★. Power saving: Automatic shutdown when no monitoring. 

 H.S.Code: 90189090

 Net Weight: 0.13kg

 Gross Weight: 0.8kg

Package Size: 210×160×70mm

 Color: Yellow, Blue 

 Standard Accessories: one SPO2 adult finger probe

Technical Specification:

General

 Monitoring Parameters: SpO2、PR、Pulse strength

Signal sockets: SpO2 socket

Display Screen    

 Type: Red LED

 Display area: 56mm×36mm

 Size: 120mm×63mm×32mm

 Weight: 125g(not include sensor and battery) 

Features

 Audio and visual alarm: The speaker sounds the alarm and the parameter’s measurement flashes on the screen.

 Prompting: Indicator of battery capacity.

 Power saving: The monitor will automatic shut down when no monitoring for 3 minutes.

Electrical Specifications

Working Voltage: D.C. 2.8V~D.C. 5.4V.

Battery 

 Type: Three common 1.5V AA alkaline or rechargeable batteries  

 Run Time: 7-hour continuous operation with a new battery (environment

temperature is 25℃)

Environment

Temperature

 Operation: 0 ~ 45 (°C)   

 Transportation and storage: -20 ~ 60 (°C)

Humidity

 Operation: 15% - 95 %( noncondensing)  

 Transportation and storage: 10% - 95 % (noncondensing)  

Altitude

 Operation: 86 KPa ~106 KPa

 Transportation and storage: 50KPa~106 KPa

Parameter Specifications

SPO2

 Patient: Adult, pediatric, infant, neonate, animal

 Range: 35 - 99%

 Resolution: 1 %

 Accuracy: ±2 % (70% - 99%), unspecified (0% - 69%)

PR

 Range: 30~250bpm

 Resolution: 1bpm

 Accuracy: ±2bpm   