GPS

DGC-I HANDHELD MULTI-PARAMETER MONITOR

★. Five Parameters: ECG, NIBP, SPO2 ,TEMP ,HR/PR  

★. Two Waveforms: ECG waveform and SPO2 waveform (PLETH).

Features:

★. Utility PC software package for long time monitoring,web transferring  

★. History data uploading/review/printing/storage

★. Friendly user interfaces,touch screen,bright color LCD real time display

★. Display the measurements and waveform at the same time

★. The ECG waveform and PLETH being switched freely

★. Storage and review trend data up to 3888 groups

★. Palm, compact ,easy for carrying and operation

★. Built-in rechargeable Li-ion battery

★. Low battery power indicator

★. Automatic shut down,the shut down time adjustable from 2-15min

Standard Packing List:

3-lead ECG cable ( include surface TEMP probe)

 Adult finger SPO2 sensor

 Adult NIBP cuff

 NIBP extension tube

PC software CD

 USB cable

 AC power adaptor

 Wall bracket  

 H.S.Code: 90181930  

 Dimension: 154×128×40mm  

 Net Weight: 480g  

 Gross Weight: 2.2kg  

 Package Size: 300X210X90mm

 Color: Yellow and White

Technical Specification:

(1) General Specifications

 Display Specification    

 Measurement Display: color LCD/touch screen    

 Power Supply Indicator: Dual-color LED (red/green)  

 Beat indicator: Dual-color LED (red/green)  

(2) Power Supply  

AC Power Adaptor    

 Input: 100~240VAC, 50/60Hz, 0.7A    

 Output: DC 4.2V, 800mA  

Battery 

 Li-ion rechargeable battery, 3.7V, 5AHr  

 Operating Time: 5 hours with full battery capacity.  

(Condition: 25℃, NIBP working period is 10 minutes)

 Recharge Time: 10 hours.

(3) Environment  

Temperature  

 Working: 0 ~ 45 (°C)    

 Transport and Storage: -25~ 85 (°C)  

Humidity  

 Working: 30% ~85 %

 Transport and Storage: 30% ~95 % (no condensation)

(4) Monitoring Parameter  

ECG

 Input: 3- lead ECG cable, standard AAMI cable connector  

 Lead Selection: I,II,III  

 Gain Selection: ×1,×2,×4,×8  

Sweep Speed: 12.5, 25, and 50 mm/s  

 Frequency Response: None filter: 0.05~100Hz (-3dB), filter: 0.5~40Hz (-3dB)  

 Calibration Signal: 1mV square wave 

 Protection: Against electrosurgical interference and defibrillation

Heart Rate

 Measuring Range: 30~250 bpm  

 Renew Time: 1 time per beat  

 Accuracy: ±1 bpm    

 Alarm Mode: Audible and visual alarm  

SPO2

 SPO2%   

 Measuring Range: 0~100%    

 Resolution: 1 %    

 Accuracy: ±2 %(70%~99%), unspecified (0%~69%)  

Pulse Rate

Measuring and Alarm Range: 0~250bpm  

 Resolution: 1bpm  

 Accuracy: ±1bpm

NIBP

 Method: Oscillometric    

 Mode: Manual, Auto, STAT  

Measuring Interval in AUTO Mode: 1, 2, 3, 5, 10, 15, 20, 30, 60 (min)  

 Measuring Period in STAT Mode: 5 min  

Cuff Pressure Range: 0~300 (mmHg) 

 Pulse Rate Range: 30~250 (bpm)  

 Measurement Rage: Adult 30~255mmHg, Pediatric 30~160mmHg  

Over-pressure Protection: Adult 300mmHg, Pediatric 220mmHg  

Pressure Resolution : 1mmHg  

 The SYSTOLIC and DIASTOLIC values measured using this device shall be equal to those measured by trained medical personnel using cuff and stethoscope. The MEAN value measured using this device is equal to that measured using blood pressure measurement device.     

Temperature

 Input: Body surface thermal-sensitive resistor temperature sensor

 Measuring Range: 0~50℃

 Accuracy: ±0.1℃

(5) Storage and Review   

 Maximum Data Quantity: 3888 groups

 Maximum Trend Time: Continuous 129 hours

    

                       Storage Screen                                        Monitor Screen

    

                        Setup Screen                             Special Software-Guardian 2009

                                                                                      for DGC-I New Style

© 2009-2019 DRAGON GROUP CHINA

                               THE BEST LUCK CORNER